Mira Family of Modelocked Ti:Sapphire Oscillators
| The Mira 900 Modelocked Titanium:Sapphire (Ti:S) laser system brings simplicity and flexibility together in an innovative design to produce a powerful and versatile system that has proven itself as the preferred ultrafast laser in thse throughout the world. |
 |
KLM - Simple, Reliable Modelocking
Modelocking is accomplished based on the widely used and reliable KLM (Kerr Lens Modelocking) technqiue. This extremely simple, all-solid-state approach is analogous to modelocking by a fast saturable absorber.
Group Velocity Dispersion (GVD) in the Mira Optima 900-F (the femtosecond cavity version of the Mira) is compensated by a simple 2-prism sequence in the cavity. Easy adjustment of these prisms balances self-phase modulation in the Ti:S rod with the GVD compensation, resulting in shortened and stabilised "soliton-like" pulses. In the Mira Optima 900-P (picosecond version) a Gires-Tournois Interferometer (GTI) conveniently suppliers the compensation.
Easy-to-use Performance
The combination of Coherent's successful DPSS 532nm laser as the pump source (5W, 8W, 10W or 18W pump power) and the Mira oscillator with the Optima system results in a laser that stands alone as the market leader in terms of ease-of-use, performance and reliability. In particular, Verdi pumping ensures the highest stability and lowest noise ultrafast performance commercially available. Mira may also be pumped by Coherent Inc's argon ion laser systems.
X-Wave Optics
Operation is greatly simplified by the use of Coherent Inc's X-Wave optics. No optics changes are required as this single optics set provides high-performance wavelength coverage from 700nm-1000nm. This feature is standard on all models.
 
Optima System
Ease of use and reliability are enhanced by the Optima system used on the Mira Optima 900 lasers. This is a series of on-board detectors whose signals are diagnosed by the laser control unit, facilitating daily operation, performance optimisation and alignment procedures.
Power/flourescence levels, CW content of the beam, pulse train and resonator humidity levels, are all displayed by the Optima system. Furthermore, automatic starting and picosecond wavelength tuning are made possible. Only Coherent inc offers the benefits of the Optima system on ultrafast lasers.
Models
Models are available in femtosecond, picosecond and dual (femto/pico) configurations. All Mira 900 models have significant configuration commonality, along with the same fundamental platform and pump lasers, ensuring the simplest upgrades and conversions.
Mira HP
The Mira HP is the world's most powerful, commercial ultrafast Ti:Sapphire oscillator. Conservatively specified at >3.3W, the Mira HP typically delivers power in excess of 4W at 800nm. Designed to be pumped by the Coherent Verdi 18W laser, this ultrafast laser delivers high power across the entire Ti:Sapphire tuning curve. The Mira HP is the latest member of the Mira family of award winning ultrafast Ti:Sapphire oscillators. These versatile lasers combine high reliability and a user-friendly open architecture with the Optima suite of on-board diagnostics.
Features include:
-
Simple, stable KLM modelocking - ease of use and reliability
-
GVD prism compensation - soliton-like, nearly transform-limited pulses
-
X-Wave optics - broadband, single optics set tuning 700-1000nm
-
Optima control and diagnostics package
-
Verdi V18 pumping (18W at 532nm) - diode-pumped stability, reliability
-
Unique resonator design - ease of use, flexibility and stability
-
Purgeable enclosure - reliability, full wavelength coverage
-
Integrated pump steering optics - ease of beam alignment
-
Optional pumping beam directions - layout convenience
-
Auxiliary CW cavity for ease of alignment or configuration change
-
Full range of ultrafast accessories
Mira Optima 900-F
The Mira Optima 900-F is the femtosecond version of the Mira product line and features the convenience of the Optima system.
Mira 900-F Tuning Curve
Features include:
-
Simple, stable KLM modelocking - ease of use and reliability
-
GVD prism compensation and GTi compensation - soliton-like, nearly transform-limited pulses
-
X-Wave optics - broadband, single optics set tuning 700-1000nm
-
Verdi pumping (5W, 8W, 10W or 18W at 532nm) - diode-pumped stability, reliability
-
Optima control and diagnostics package
-
Unique resonator design - ease of use, flexibility and stability
-
Purgeable enclosure - reliability, full wavelength coverage
-
Integrated pump steering optics - ease of beam alignment
-
Optional pumping beam directions - layout convenience
-
Argon ion pump laser configurations - flexibility
-
Auxiliary CW cavity for ease of alignment or configuration change
-
Frequency conversion for blue/UV operation
-
Short-pulse short
-
Pulse-picker - repetition-rate control option
-
OPO, regenerative amplifier and OPA options
Mira Optima 900-P
The Mira Optima 900-P is the picosecond version of the Mira product line and features the convenience of the Optima system. In particular, Optima allows continuous, skip-free tuning in picosecond operation. GVD (Group Velocity Dispersion) control is conveniently achieved by the use of a Gires-Tournois Interferometer (GTI).
Mira 900-P Tuning Curve
-
Simple, stable KLM modelocking - ease of use and reliability
-
GVD prism compensation and GTi compensation - soliton-like, nearly transform-limited pulses
-
X-Wave optics - broadband, single optics set tuning 700-1000nm
-
Verdi pumping (5W, 8W, 10W or 18W at 532nm) - diode-pumped stability, reliability
-
Optima control and diagnostics package
-
Unique resonator design - ease of use, flexibility and stability
-
Purgeable enclosure - reliability, full wavelength coverage
-
Integrated pump steering optics - ease of beam alignment
-
Optional pumping beam directions - layout convenience
-
Argon ion pump laser configurations - flexibility
-
Auxiliary CW cavity for ease of alignment or configuration change
-
Frequency conversion for blue/UV operation
-
Short-pulse short
-
Pulse-picker - repetition-rate control option
-
OPO, regenerative amplifier and OPA options
Mira Optima 900-D
The Mira Optima 900-D (Dual) offers maximum versatility in terms of application potential, combining the features of the 900-F and 900-P in a simple dual-oscialltor design. The system configuration is conveniently changed from femto to pico and vice-versa. All necessary optics are provided.
-
Simple, stable KLM modelocking - ease of use and reliability
-
GVD prism compensation and GTi compensation - soliton-like, nearly transform-limited pulses
-
X-Wave optics - broadband, single optics set tuning 700-1000nm
-
Verdi pumping (5W, 8W, 10W or 18W at 532nm) - diode-pumped stability, reliability
-
Optima control and diagnostics package
-
Unique resonator design - ease of use, flexibility and stability
-
Purgeable enclosure - reliability, full wavelength coverage
-
Integrated pump steering optics - ease of beam alignment
-
Optional pumping beam directions - layout convenience
-
Argon ion pump laser configurations - flexibility
-
Auxiliary CW cavity for ease of alignment or configuration change
-
Frequency conversion for blue/UV operation
-
Short-pulse short
-
Pulse-picker - repetition-rate control option
-
OPO, regenerative amplifier and OPA options
Mira-S
The Mira 900-S provides a more economical solution for femtosecond pulse generation. The 900-S offers many of the features and benefits of the Mira family, including Verdi pumping and X-Wave optics. The Optima system is not included in the 900-S, but a manually-activated started for modelocking is provided with the system.
-
Simple, stable KLM modelocking - ease of use and reliability
-
GVD prism compensation and GTi compensation - soliton-like, nearly transform-limited pulses
-
X-Wave optics - broadband, single optics set tuning 700-1000nm
-
Verdi pumping (5W, 8W, 10W or 18W at 532nm) - diode-pumped stability, reliability
-
Unique resonator design - ease of use, flexibility and stability
-
Purgeable enclosure - reliability, full wavelength coverage
-
Integrated pump steering optics - ease of beam alignment
-
Optional pumping beam directions - layout convenience
-
Argon ion pump laser configurations - flexibility
-
Auxiliary CW cavity for ease of alignment or configuration change
-
Frequency conversion for blue/UV operation
-
Short-pulse short
-
Pulse-picker - repetition-rate control option
-
OPO option |
|